Frontier® Acid Digestion™

Frontier® Acid Digestion™
  • Tri-wall construction and chain-and-sprocket sash system for maximum robustness
  • u-PVC exhaust collar ensures chemical resistance
  • Polycarbonate sash to prevent etching
  • Esco’s Creep down Mechanism: When sash is raised above 457 mm (18"), sash will automatically and gently fall back to the safe level.
  • Baffle system rapidly draws contaminant out of the fume hood and thermal heat relief is achieved.
  • Service fixtures include: 1 x water fixture, 1 x gas fixture, 4 single electrical outlets with splash proof cover and 1 PP drip cup.

Guide to Models

EFQ- _ D _ _ W - _

External Width

Code

Internal Depth

Code

Internal Liner

Code

Sash Type

Code

Color

Code

Electrical

Code

1220 mm (48.0”)

4U

900 mm (35.4”)

D

u-PVC

C

Vertical

V

Esco White

W

220-240 VAC, 50-60 Hz, 1Ø

8

1525 mm (60.0”)

5U

 

 

Polypropylene

P

Combination

C

 

 

110-120 VAC, 50-60 Hz, 1Ø

9

1830 mm (72.0”)

6U

 

 

 

 

 

 

 

 

 

 

2440 mm (96.0”)

8U

 

 

 

 

 

 

 

 

 

 

 

Model

220-240 VAC,
50-60 Hz, 1Ø

EFQ-4UDCCW-8

EFQ-4UDCVW-8

EFQ-5UDCCW-8

EFQ-5UDCVW-8

EFQ-6UDCCW-8

EFQ-6UDCVW-8

EFQ-8UDCCW-8

EFQ-8UDCVW-8

EFQ-4UDPCW-8

EFQ-4UDPVW-8

EFQ-5UDPCW-8

EFQ-5UDPVW-8

EFQ-6UDPCW-8

EFQ-6UDPVW-8

EFQ-8UDPCW-8

EFQ-8UDPVW-8

110-120 VAC,
 50-60 Hz, 1Ø

EFQ-4UDCCW-9

EFQ-4UDCVW-9

EFQ-5UDCCW-9

EFQ-5UDCVW-9

EFQ-6UDCCW-9

EFQ-6UDCVW-9

EFQ-8UDCCW-9

EFQ-8UDCVW-9

EFQ-4UDPCW-9

EFQ-4UDPVW-9

EFQ-5UDPCW-9

EFQ-5UDPVW-9

EFQ-6UDPCW-9

EFQ-6UDPVW-9

EFQ-8UDPCW-9

EFQ-8UDPVW-9

Nominal Size

1.2 meters (4’)

1.5 meters (5’)

1.8 meters (6’)

2.4 meters (8’)

External Dimensions
 (W x D x H)

1220 x 900 x 1644 mm

(48.0” x 35.4”x 64.7”)

1525 x 900 x 1644 mm

(60.0” x 35.4”x 64.7”)

1830 x 900 x 1644 mm

(72.0” x 35.4”x 64.7”)

2440 x 900 x 1644 mm

(96.0” x 35.4”x 64.7”)

Internal Work Area (W x D x H)

996 x 672 x 1240 mm

(39.2” x 26.5”x 48.8”)

1301 x 672 x 1240 mm

(51.2” x 26.5”x 48.8”)

1606 x 672 x 1240 mm

(63.2” x 26.5”x 48.8”)

2216 x 672 x 1240 mm

(87.2” x 26.5”x 48.8”)

Maximum sash Opening

740 mm (29.1”)

Face Velocity

Sash Opening

Exhaust volume / static pressure required

0.3 m/s
(60 fpm)

457 mm
(18.0”)

542 cmh at 7 Pa

319 cfm at 0.03" WG

705 cmh at 9 Pa

415 cfm at 0.04" WG

867 cmh at 12 Pa

510 cfm at 0.05" WG

1192 cmh at 8 Pa

702 cfm at 0.03" WG

0.4 m/s
(80 fpm)

723 cmh at 13 Pa

425 cfm at 0.05" WG

939 cmh at 17 Pa

553 cfm at 0.07" WG

1156 cmh at 20 Pa

681 cfm at 0.09" WG

1590 cmh at 14 Pa

936 cfm at 0.06" WG

0.5 m/s
(100 fpm)

903 cmh at 20 Pa

532 cfm at 0.09" WG

1174 cmh at 26 Pa

691 cfm at 0.11" WG

1445 cmh at 28 Pa

851 cfm at 0.12" WG

1987 cmh at 22 Pa

1170 cfm at 0.09" WG

0.3 m/s
(60 fpm)

736 mm
(29.0”)

642 cmh at 13 Pa

378 cfm at 0.06" WG

834 cmh at 17 Pa

491 cfm at 0.07" WG

1027 cmh at 22 Pa

605 cfm at 0.09" WG

1412 cmh at 14 Pa

831 cfm at 0.06" WG

0.4 m/s
(80 fpm)

856 cmh at 23 Pa

504 cfm at 0.10" WG

1113 cmh at 30 Pa

655 cfm at 0.13" WG

1369 cmh at 39 Pa

806 cfm at 0.17" WG

1883 cmh at 25 Pa

1108 cfm at 0.11" WG

0.5 m/s
(100 fpm)

1070 cmh at 36 Pa

630 cfm at 0.15" WG

1391 cmh at 47 Pa

819 cfm at 0.20" WG

1712 cmh at 61 Pa

1008 cfm at 0.26" WG

2354 cmh at 40 Pa

1385 cfm at 0.17" WG

Exhaust Outlet Diameter

305 mm (12.0”)

Number of exhaust Collars

1

2

  • All Frontier Acela Fume Hoods include 1 water and 1 gas Enhanz™ remote-controlled fixture plumbed to the top of the hood and 2 duplex outlets standard. (Specify outlet code when ordering)
  • Base Cabinet (EBA)
    • Optional Hood circuit protection (EBA-8)
       
  • Ventilation Kit (VK-EBA)
    • Ventilates base cabinet utilizing the hood exhaust system.
    • Field-installed.
       
  • Filler Panel (FP-EBAD)
    • One set of filler panels required per continuous row of hoods.
    • Field-installed
       
  • Special Flammable and Corrosive storage cabinets (ESF-F/C)