Frontier PPH® Fume Hood

Frontier PPH® Fume Hood
1. Full polypropylene(PP) interior and exterior
  • Metal-free: suitable for sensitive work such as trace metal analysis
  • Wall: stress-relieved, acid-resistant, non-corrosive, bend-resistant
  • Work-top: excellent acid and alkali erosion endurance, high heat and decoloration resistance

2. Operates safely at 0.5m/s at 18" (457mm) or full open sash position constant volume concept eliminates the need forcomplex VAV systems.

3. Optional Sentinel XL Airflow Monitor accurately monitors face velocity and alarms when inflow decreases to unsafe levels.

4. The tapered PP exhaust collar reduces airflow turbulence, static pressure loss, noise level and increases face velocity uniformity.

5. Perforated Baffles: improve capture efficiency.

6. Air Foil: reduce turbulence and eliminate backflow.

7. Sturcture design
  • Sash: tempered safety polycarbonate, can be fully opened and set in any working position
  • Base Cabinet: one shelf fixed in the left cabinet to prevent liquid leakage

8. Remote control water fixture and 1 gas fixture are factory installed and plumbed, Up to 4 services may be factory or field installed.

9. Perfect Pitch Profile: 5° Angled Front, this gentle pitch of the hood profile enhances visibility and maximizes operator comfort.

10. Removable structure for easy access: baffles, panel on internal side wall and rear access panel.

General Specifications, Frontier® PPH™ Series
PPH-_DP_W-_
External Width Code External Depth Code Internal Liner Code Sash Type Code Color Code Electrical Code
1220 mm (48.0") 4U 900 mm (35.4") M Esco PP P Vertical V Esco White W 220-240 VAC, 50/60 Hz 8
1525 mm (60.0") 5U         Combination C     110-120 VAC, 50/60 Hz 9
1800 mm (70.8") 6U                    
2400 mm (94.5") 8U                    
General Specifications, Frontier® PPH™ Series
Model 220-240 VAC,
50/60 Hz, 1ø
PPH-4UDPVW-8 PPH-5UDPVW-8 PPH-6UDPVW-8 PPH-8UDPVW-8
PPH-4UDPCW-8 PPH-5UDPCW-8 PPH-6UDPCW-8 PPH-8UDPCW-8
110-120 VAC,
50/60 Hz, 1ø
PPH-4UDPVW-9 PPH-5UDPVW-9 PPH-6UDPVW-9 PPH-8UDPVW-9
PPH-4UDPCW-9 PPH-5UDPCW-9 PPH-6UDPCW-9 PPH-8UDPCW-9
Nominal Size 1.2 meter (4') 1.5 meter (5') 1.8 meter (6') 2.4 meter (8')
External Dimensions
(W x D x H)
Fume Hood unit only 1200 x 900 x 1500 mm
(47.2" x 35.4" x 59.1")
1500 x 900 x 1500 mm
(59.0" x 35.4" x 59.1")
1800 x 900 x 1500 mm
(70.8" x 35.4" x 59.1")
2400 x 900 x 1500 mm
(94.5" x 35.4" x 59.1")
With Exhaust Collar 1200 x 900 x 1681 mm
(47.2" x 35.4" x 66.2")
1500 x 900 x 1681 mm
(59.0" x 35.4" x 66.2")
1800 x 900 x 1681 mm
(70.8" x 35.4" x 66.2")
2400 x 900 x 1681 mm
(94.5" x 35.4" x 66.2")
With Fully-opened Sash 1200 x 900 x 1879 mm
(47.2" x 35.4" x 74.0")
1500 x 900 x 1879 mm
(59.0" x 35.4" x 74.0")
1800 x 900 x 1879 mm
(70.8" x 35.4" x 74.0")
2400 x 900 x 1879 mm
(94.5" x 35.4" x 74.0")
Internal Dimensions
(W x D x H)
980 x 665 x 1200 mm
(38.6" x 26.1" x 47.2")
1280 x 665 x 1200 mm
(50.4" x 26.1" x 47.2")
1580 x 665 x 1200 mm
(62.2" x 26.1" x 47.2")
2180 x 665 x 1200 mm
(85.8" x 26.1" x 47.2")
Construction Main Body Polypropylene
Internal Liner
Worktop (default)
Sash Specification Sash Material Polycarbonate
Sash Configuration Vertical or Combination
Sloping 5° Slope
Maximum Sash Opening
(two vertical sash opening)
790 mm (31.1")

  • Sentinel XL airflow monitor (discussed above).
  • Additional service fixtures and electrical outlets.
  • Exhaust blower